CVE-2014-5406
Hospira LifeCare PCA Infusion System
- State
- Published
- Severity
- 7.6 High
- Published
- Jul 6, 2015
- Updated
- Nov 3, 2025
- Reserved
- Aug 22, 2014
Description
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 2.0 | 7.6 High | AV:N/AC:H/Au:N/C:C/I:C/A:C | icscert |
Weaknesses
Affected Products
| Product | Versions |
|---|---|
LifeCare PCA Infusion SystemHospira |
|
References
- https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.