CVE-2014-5431

State
Published
Severity
Not scored
Published
Mar 26, 2019
Updated
Aug 6, 2024
Reserved
Aug 22, 2014

Read the record on cve.org

Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Scores

The record has no CVSS score.

Weaknesses

  • CWE-259 Hard-coded password CWE-259

Affected Products

ProductVersions
SIGMA Spectrum Infusion SystemBaxter
  • 6.05 (model 35700BAX) with wireless battery module (WBM) version 16Affected

References

Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.