CVE-2014-5433
- State
- Published
- Severity
- Not scored
- Published
- Mar 26, 2019
- Updated
- Aug 6, 2024
- Reserved
- Aug 22, 2014
Description
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Scores
The record has no CVSS score.
Weaknesses
- CWE-312 Cleartext storage of sensitive information CWE-312
Affected Products
| Product | Versions |
|---|---|
SIGMA Spectrum Infusion SystemBaxter |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.