CVE-2014-5433

State
Published
Severity
Not scored
Published
Mar 26, 2019
Updated
Aug 6, 2024
Reserved
Aug 22, 2014

Read the record on cve.org

Description

An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Scores

The record has no CVSS score.

Weaknesses

  • CWE-312 Cleartext storage of sensitive information CWE-312

Affected Products

ProductVersions
SIGMA Spectrum Infusion SystemBaxter
  • 6.05 (model 35700BAX) with wireless battery module (WBM) version 16Affected

References

Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.