CVE-2015-1012
- State
- Published
- Severity
- Not scored
- Published
- Mar 25, 2019
- Updated
- Aug 6, 2024
- Reserved
- Jan 10, 2015
Description
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Scores
The record has no CVSS score.
Weaknesses
- CWE-312 Cleartext storage of sensitive information CWE-312
Affected Products
| Product | Versions |
|---|---|
LifeCare PCA Infusion SystemHospira |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.