CVE-2015-3952

State
Published
Severity
Not scored
Published
Mar 25, 2019
Updated
Aug 6, 2024
Reserved
May 12, 2015

Read the record on cve.org

Description

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Scores

The record has no CVSS score.

Weaknesses

  • CWE-312 Cleartext storage of sensitive information CWE-312

Affected Products

ProductVersions
Plum A+ Infusion SystemHospira
  • <= 13.4Affected
Plum A+3 Infusion SystemHospira
  • <= 13.6Affected
Symbiq Infusion SystemHospira
  • <= 3.13Affected