CVE-2015-3953

State
Published
Severity
Not scored
Published
Mar 25, 2019
Updated
Aug 6, 2024
Reserved
May 12, 2015

Read the record on cve.org

Description

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Scores

The record has no CVSS score.

Weaknesses

  • CWE-259 Use of hard-coded password CWE-259

Affected Products

ProductVersions
Plum A+ Infusion SystemHospira
  • <= 13.4Affected
Plum A+3 Infusion SystemHospira
  • <= 13.6Affected
Symbiq Infusion SystemHospira
  • <= 3.13Affected

References

Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.