CVE-2017-12716
- State
- Published
- Severity
- Not scored
- Published
- Apr 25, 2018
- Updated
- Sep 17, 2024
- Reserved
- Aug 9, 2017
Description
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Scores
The record has no CVSS score.
Weaknesses
- CWE-311 Missing encryption of sensitive data CWE-311
Affected Products
| Product | Versions |
|---|---|
Accent and AnthemAbbott Laboratories |
|
References