CVE-2017-9658

State
Published
Severity
Not scored
Published
Apr 30, 2018
Updated
Sep 16, 2024
Reserved
Jun 14, 2017

Read the record on cve.org

Description

Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Scores

The record has no CVSS score.

Weaknesses

  • CWE-755 Improper handling of exceptional conditions CWE-755

Affected Products

ProductVersions
IntelliVue MX40 Patient Worn MonitorPhilips
  • IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18Affected

References

Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.