CVE-2018-8870
Medtronic MyCareLink Patient Monitor Use of Hard-coded Password
- State
- Published
- Severity
- 6.4 Medium
- Published
- Jul 2, 2018
- Updated
- May 22, 2025
- Reserved
- Mar 20, 2018
Description
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 3.1 | 6.4 Medium | CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H | icscert |
Weaknesses
Affected Products
| Product | Versions |
|---|---|
24950 MyCareLink MonitorMedtronic |
|
24952 MyCareLink MonitorMedtronic |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.