CVE-2019-18248
- State
- Published
- Severity
- Not scored
- Published
- Jun 29, 2020
- Updated
- Aug 5, 2024
- Reserved
- Oct 22, 2019
Description
BIOTRONIK CardioMessenger II, The affected products transmit credentials in clear-text prior to switching to an encrypted communication channel. An attacker can disclose the product’s client credentials for connecting to the BIOTRONIK Remote Communication infrastructure.
Scores
The record has no CVSS score.
Weaknesses
- CWE-319 CLEARTEXT TRANSMISSION OF SENSITIVE INFORMATION CWE-319
Affected Products
| Product | Versions |
|---|---|
BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSMn/a |
|
References