CVE-2019-25721

Dräger Infinity M300 VG2.3.1 Network-Based Denial of Service

State
Published
Severity
7.1 High
Published
Jun 2, 2026
Updated
Jun 3, 2026
Reserved
Jun 2, 2026

Read the record on cve.org

Description

Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.

Scores

CVSSScoreVectorSource
4.07.1 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:NVulnCheck
3.16.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:HVulnCheck

Weaknesses

  • CWE-400 Uncontrolled Resource Consumption

Affected Products

ProductVersions
Infinity M300Dräger
  • VG2.3.1 and earlierAffected
  • VG2.3.2Not affected