CVE-2019-25722

Dräger SC Monitoring Devices Hard-coded Credentials and DoS

State
Published
Severity
7.2 High
Published
Jun 2, 2026
Updated
Jun 3, 2026
Reserved
Jun 2, 2026

Read the record on cve.org

Description

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.

Scores

CVSSScoreVectorSource
4.07.2 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:NVulnCheck
3.17.6 HighCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:HVulnCheck

Weaknesses

  • CWE-798 Use of Hard-coded Credentials

Affected Products

ProductVersions
SC 6002XLDräger
  • SC 6002XLAffected
SC6802XLDräger
  • SC6802XLAffected
SC 7000Dräger
  • SC 7000Affected
SC8000Dräger
  • SC8000Affected
SC90000 XLDräger
  • SC90000 XLAffected

References

Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.