CVE-2019-25723
Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface
- State
- Published
- Severity
- 6.3 Medium
- Published
- Jun 2, 2026
- Updated
- Jun 3, 2026
- Reserved
- Jun 2, 2026
Description
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 4.0 | 6.3 Medium | CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L | VulnCheck |
| 3.1 | 4.0 Medium | CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L | VulnCheck |
Weaknesses
- CWE-1286 Improper Validation of Syntactic Correctness of Input
Affected Products
| Product | Versions |
|---|---|
Perseus A500Dräger |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.