CVE-2019-25724

Dräger Infinity M300 VG2.x Network-Based Denial of Service

State
Published
Severity
7.1 High
Published
Jun 2, 2026
Updated
Jun 3, 2026
Reserved
Jun 2, 2026

Read the record on cve.org

Description

Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

Scores

CVSSScoreVectorSource
4.07.1 HighCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:NVulnCheck
3.16.5 MediumCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:HVulnCheck

Weaknesses

  • CWE-400 Uncontrolled Resource Consumption

Affected Products

ProductVersions
Infinity M300Dräger
  • VG2.x through VG2.3.1Affected
  • VG3.0Not affected

References