CVE-2019-6538

Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

State
Published
Severity
9.3 Critical
Published
Mar 25, 2019
Updated
May 22, 2025
Reserved
Jan 22, 2019

Read the record on cve.org

Description

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Scores

CVSSScoreVectorSource
3.19.3 CriticalCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:Hicscert

Weaknesses

Affected Products

ProductVersions
Conexus Radio Frequency Telemetry ProtocolMedtronic
  • All versionsAffected
MyCareLink MonitorMedtronic
  • 24950Affected
  • 24952Affected
CareLink MonitorMedtronic
  • 2490CAffected
CareLink 2090 ProgrammerMedtronic
  • All versionsAffected
Amplia CRT-DMedtronic
  • All versionsAffected
Claria CRT-DMedtronic
  • All versionsAffected
Compia CRT-DMedtronic
  • All versionsAffected
Concerto CRT-DMedtronic
  • All versionsAffected
Concerto II CRT-DMedtronic
  • All versionsAffected
Consulta CRT-DMedtronic
  • All versionsAffected
Evera ICDMedtronic
  • All versionsAffected
Maximo II CRT-DMedtronic
  • All versionsAffected
Maximo II ICDMedtronic
  • All versionsAffected
Mirro ICDMedtronic
  • All versionsAffected
Nayamed ND ICDMedtronic
  • All versionsAffected
Primo ICDMedtronic
  • All versionsAffected
Protecta ICD, Protecta CRT-DMedtronic
  • All versionsAffected
Secura ICDMedtronic
  • All versionsAffected
Virtuoso ICDMedtronic
  • All versionsAffected
Virtuoso II ICDMedtronic
  • All versionsAffected
Visia AF ICDMedtronic
  • All versionsAffected
Viva CRT-DMedtronic
  • All versionsAffected
Brava CRT-DMedtronic
  • All versionsAffected
Mirro MRI ICDMedtronic
  • All versionsAffected

References