CVE-2020-12036
- State
- Published
- Severity
- Not scored
- Published
- Jun 29, 2020
- Updated
- Aug 4, 2024
- Reserved
- Apr 21, 2020
Description
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
Scores
The record has no CVSS score.
Weaknesses
- CWE-319 CLEARTEXT TRANSMISSION OF SENSITIVE INFORMATION CWE-319
Affected Products
| Product | Versions |
|---|---|
Baxter PrismaFlex and PrisMaxn/a |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.