CVE-2020-12040
- State
- Published
- Severity
- Not scored
- Published
- Jun 29, 2020
- Updated
- Aug 4, 2024
- Reserved
- Apr 21, 2020
Description
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Scores
The record has no CVSS score.
Weaknesses
Affected Products
| Product | Versions |
|---|---|
Baxter Sigma Spectrum Infusion Pumpsn/a |
|