CVE-2020-12041
- State
- Published
- Severity
- Not scored
- Published
- Jun 29, 2020
- Updated
- Aug 4, 2024
- Reserved
- Apr 21, 2020
Description
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.
Scores
The record has no CVSS score.
Weaknesses
- CWE-732 INCORRECT PERMISSION ASSIGNMENT FOR CRITICAL RESOURCE CWE-732
Affected Products
| Product | Versions |
|---|---|
Baxter Sigma Spectrum Infusion Pumpsn/a |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.