CVE-2021-23236

Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption

State
Published
Severity
7.5 High
Published
Jan 21, 2022
Updated
Apr 16, 2025
Reserved
Nov 30, 2021

Read the record on cve.org

Description

Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system.

Scores

CVSSScoreVectorSource
3.17.5 HighCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:Hicscert

Weaknesses

  • CWE-400 Uncontrolled Resource Consumption

Affected Products

ProductVersions
Agilia Link+Fresenius Kabi
  • before 3.0Affected

References

Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.