CVE-2022-36385
Contec Health CMS8000
- State
- Published
- Severity
- 6.8 Medium
- Published
- Sep 13, 2022
- Updated
- Apr 16, 2025
- Reserved
- Aug 29, 2022
Description
A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 3.1 | 6.8 Medium | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H | icscert |
Weaknesses
- CWE-284 Improper Access Control
Affected Products
| Product | Versions |
|---|---|
CMS8000 CONTEC ICU CCU Vital Signs Patient MonitorContec Health |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.