CVE-2022-38100
Contec Health CMS8000
- State
- Published
- Severity
- 7.5 High
- Published
- Sep 13, 2022
- Updated
- Apr 16, 2025
- Reserved
- Aug 29, 2022
Description
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 3.1 | 7.5 High | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H | icscert |
Weaknesses
- CWE-400 Uncontrolled Resource Consumption
Affected Products
| Product | Versions |
|---|---|
CMS8000 CONTEC ICU CCU Vital Signs Patient MonitorContec Health |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.