CVE-2022-4992

Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering

State
Published
Severity
8.8 High
Published
Jun 2, 2026
Updated
Jun 5, 2026
Reserved
Jun 2, 2026

Read the record on cve.org

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

Scores

CVSSScoreVectorSource
4.08.8 HighCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:NVulnCheck
3.18.6 HighCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:HVulnCheck

Weaknesses

  • CWE-345 Insufficient Verification of Data Authenticity

Affected Products

ProductVersions
Infinity Acute Care SystemDräger
  • before VG4.2Affected
  • before VG4.1.1Affected
  • before VG4.0.3Affected
Standalone Infinity M540 patient monitorDräger
  • before VG4.2Affected
  • before VG4.1.1Affected

References

Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.