CVE-2023-25931

Medtronic Micro Clinician & InterStim X Clinician App Password Reset Issue

State
Published
Severity
6.4 Medium
Published
Mar 1, 2023
Updated
Mar 7, 2025
Reserved
Feb 16, 2023

Read the record on cve.org

Description

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Scores

CVSSScoreVectorSource
3.16.4 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:HMedtronic

Weaknesses

Affected Products

ProductVersions
InsterStim ApplicationsMedtronic
  • Micro Clinician to before A51200Affected
  • InterStim X Clinician to before A51300Affected