CVE-2024-48966
Life2000 service tools for test and calibration do not support user authentication
- State
- Published
- Severity
- 10.0 Critical
- Published
- Nov 14, 2024
- Updated
- Nov 15, 2024
- Reserved
- Oct 10, 2024
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 3.1 | 10.0 Critical | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H | Baxter |
Weaknesses
- CWE-306 Missing Authentication for Critical Function
Affected Products
| Product | Versions |
|---|---|
Life2000 Ventilation SystemBaxter |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.