CVE-2024-48971

Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator

State
Published
Severity
9.3 Critical
Published
Nov 14, 2024
Updated
Nov 15, 2024
Reserved
Oct 10, 2024

Read the record on cve.org

Description

The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.

Scores

CVSSScoreVectorSource
3.19.3 CriticalCVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:HBaxter

Weaknesses

  • CWE-798 Use of Hard-coded Credentials

Affected Products

ProductVersions
Life2000 Ventilation SystemBaxter
  • 06.08.00.00 and priorAffected