CVE-2024-48973

Debug port on Life2000 Ventilator serial interface is enabled by default

State
Published
Severity
9.3 Critical
Published
Nov 14, 2024
Updated
Nov 18, 2024
Reserved
Oct 10, 2024

Read the record on cve.org

Description

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Scores

CVSSScoreVectorSource
3.19.3 CriticalCVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:HBaxter

Weaknesses

  • CWE-1263 Improper Physical Access Control

Affected Products

ProductVersions
Life2000 Ventilation SystemBaxter
  • 06.08.00.00 and priorAffected

References

Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.