CVE-2024-48974
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
- State
- Published
- Severity
- 9.3 Critical
- Published
- Nov 14, 2024
- Updated
- Nov 18, 2024
- Reserved
- Oct 10, 2024
Description
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 3.1 | 9.3 Critical | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H | Baxter |
Weaknesses
- CWE-494 Download of Code Without Integrity Check
Affected Products
| Product | Versions |
|---|---|
Life2000 Ventilation SystemBaxter |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.