CVE-2024-48974

Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

State
Published
Severity
9.3 Critical
Published
Nov 14, 2024
Updated
Nov 18, 2024
Reserved
Oct 10, 2024

Read the record on cve.org

Description

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.

Scores

CVSSScoreVectorSource
3.19.3 CriticalCVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:HBaxter

Weaknesses

  • CWE-494 Download of Code Without Integrity Check

Affected Products

ProductVersions
Life2000 Ventilation SystemBaxter
  • 06.08.00.00 and priorAffected

References

Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.