CVE-2024-9832

No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator

State
Published
Severity
9.3 Critical
Published
Nov 14, 2024
Updated
Nov 18, 2024
Reserved
Oct 10, 2024

Read the record on cve.org

Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Scores

CVSSScoreVectorSource
3.19.3 CriticalCVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:HBaxter

Weaknesses

  • CWE-307 Improper Restriction of Excessive Authentication Attempts

Affected Products

ProductVersions
Life2000 Ventilation SystemBaxter
  • 06.08.00.00 and priorAffected

References

Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.