CVE-2024-9832
No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
- State
- Published
- Severity
- 9.3 Critical
- Published
- Nov 14, 2024
- Updated
- Nov 18, 2024
- Reserved
- Oct 10, 2024
Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 3.1 | 9.3 Critical | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H | Baxter |
Weaknesses
- CWE-307 Improper Restriction of Excessive Authentication Attempts
Affected Products
| Product | Versions |
|---|---|
Life2000 Ventilation SystemBaxter |
|