CVE-2025-0626

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

State
Published
Severity
7.7 High
Published
Jan 30, 2025
Updated
Mar 1, 2025
Reserved
Jan 21, 2025

Read the record on cve.org

Description

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Scores

CVSSScoreVectorSource
4.07.7 HighCVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:Nicscert
3.17.5 HighCVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:Hicscert

Weaknesses

Affected Products

ProductVersions
CMS8000 Patient MonitorContec Health
  • All versionsAffected

References