CVE-2025-0683

Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

State
Published
Severity
8.2 High
Published
Jan 30, 2025
Updated
Feb 12, 2025
Reserved
Jan 23, 2025

Read the record on cve.org

Description

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Scores

CVSSScoreVectorSource
4.08.2 HighCVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:Nicscert
3.15.9 MediumCVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:Nicscert

Weaknesses

  • CWE-359 Exposure of Private Personal Information to an Unauthorized Actor

Affected Products

ProductVersions
CMS8000 Patient MonitorContec Health
  • All versionsAffected

References