CVE-2025-14095

Privilege boundary violation in Radiometer Products

State
Published
Severity
6.8 Medium
Published
Dec 17, 2025
Updated
Dec 17, 2025
Reserved
Dec 5, 2025

Read the record on cve.org

Description

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.

Other related CVE's are CVE-2025-14096 & CVE-2025-14097.

Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required configuration for Exposure:

Physical access to the analyzer is needed.

Temporary work Around:

Only authorized people can physically access the analyzer.

Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note:

CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

Scores

CVSSScoreVectorSource
3.16.8 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HRadiometer
3.15.7 MediumCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:HRadiometer

Weaknesses

  • CWE-284 Improper Access Control
  • CWE-693 — Protection Mechanism Failure

Affected Products

ProductVersions
ABL90 FLEX and ABL90 FLEX PLUS AnalyzersRadiometer Medical Aps
  • All application software versions with Windows 7, Windows XP as underlying OSAffected
  • Application software versions < 3.5MR11 with Windows 10 as underlying OSAffected
AQT90 FLEX AnalyzersRadiometer Medical Aps
  • All Application software versions <= 8.13 MR2Affected
ABL800 BASIC and ABL800 FLEX AnalyzersRadiometer Medical Aps
  • Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OSAffected
  • Application software versions < 6.20 MR2 with Windows 10 as underlying OSAffected
ABL9 AnalyzersRadiometer Medical Aps
  • Application software versions < 1.5.0Affected