CVE-2025-14096
Credential Disclosure vulnerability in Radiometer Products
- State
- Published
- Severity
- 8.4 High
- Published
- Dec 17, 2025
- Updated
- Dec 17, 2025
- Reserved
- Dec 5, 2025
Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.
Other related CVE's are CVE-2025-14095 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure:
Attacker requires physical access to the analyzer.
Temporary work Around: Only authorized people can physically access the analyzer.
Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 3.1 | 8.4 High | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H | Radiometer |
Weaknesses
Affected Products
| Product | Versions |
|---|---|
ABL90 FLEX and ABL90 FLEX PLUS AnalyzersRadiometer Medical Aps |
|
AQT90 FLEX AnalyzersRadiometer Medical Aps |
|
ABL800 BASIC and ABL800 FLEX analyzersRadiometer Medical Aps |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.