CVE-2025-14097

Remote Code Execution Vulnerability in Radiometer Products

State
Published
Severity
7.2 High
Published
Dec 17, 2025
Updated
Dec 17, 2025
Reserved
Dec 5, 2025

Read the record on cve.org

Description

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status:

Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Scores

CVSSScoreVectorSource
3.17.2 HighCVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:HRadiometer

Weaknesses

Affected Products

ProductVersions
ABL90 FLEX and ABL90 FLEX PLUS AnalyzersRadiometer Medical Aps
  • Application software versions < 3.5MR11 with Windows 7, Windows XP operating systemsAffected
  • Application software versions >= 3.5MR11 with Windows 10 operating systemNot affected
ABL800 BASIC and ABL800 FLEX AnalyzersRadiometer Medical Aps
  • Application software versions < 6.20MR2 with Windows 7, Windows XP operating systemsAffected
  • Application software versions >= 6.20MR2 with Windows 10 operating systemNot affected
AQT90 FLEX AnalyzersRadiometer Medical Aps
  • Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systemsAffected
  • Application software versions >= 8.13 MR2 with Windows 10 operating systemNot affected