CVE-2025-15653
Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation
- State
- Published
- Severity
- 7.0 High
- Published
- Jun 2, 2026
- Updated
- Jun 3, 2026
- Reserved
- Jun 2, 2026
Description
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 4.0 | 7.0 High | CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N | VulnCheck |
| 3.1 | 6.8 Medium | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H | VulnCheck |
Weaknesses
- CWE-668 Exposure of Resource to Wrong Sphere
Affected Products
| Product | Versions |
|---|---|
Zeus IEDräger |
|
Zeus RS C500Dräger |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.