CVE-2025-41690
Endress+Hauser: Proline 10 Maintenance credentials may be exposed under certain conditions
- State
- Published
- Severity
- 7.4 High
- Published
- Sep 2, 2025
- Updated
- Sep 2, 2025
- Reserved
- Apr 16, 2025
Description
A low-privileged attacker in bluetooth range may be able to access the password of a higher-privilege user (Maintenance) by viewing the device’s event log. This vulnerability could allow the Operator to authenticate as the Maintenance user, thereby gaining unauthorized access to sensitive configuration settings and the ability to modify device parameters.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 3.1 | 7.4 High | CVSS:3.1/AV:A/AC:L/PR:L/UI:R/S:U/C:H/I:H/A:H | CERTVDE |
Weaknesses
- CWE-532 Insertion of Sensitive Information into Log File
Affected Products
| Product | Versions |
|---|---|
Promag 10 with HARTEndress+Hauser |
|
Promag 10 with IO-LinkEndress+Hauser |
|
Promag 10 with ModbusEndress+Hauser |
|
Promass 10 with HARTEndress+Hauser |
|
Promass 10 with IO-LinkEndress+Hauser |
|
Promass 10 with ModbusEndress+Hauser |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.