CVE-2025-64298
Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource
- State
- Published
- Severity
- 8.6 High
- Published
- Dec 2, 2025
- Updated
- Dec 9, 2025
- Reserved
- Nov 11, 2025
Description
NMIS/BioDose V22.02 and previous version installations where the embedded Microsoft SQLServer Express is used are exposed in the Windows share accessed by clients in networked installs. By default, this directory has insecure directory paths that allow access to the SQL Server database and configuration files, which can contain sensitive data.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 4.0 | 8.6 High | CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N | icscert |
| 3.1 | 8.4 High | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H | icscert |
Weaknesses
- CWE-732 Incorrect Permission Assignment for Critical Resource
Affected Products
| Product | Versions |
|---|---|
EC2 Software NMIS BioDoseMirion Medical |
|
References
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.