CVE-2026-0611
Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting
- State
- Published
- Severity
- 9.2 Critical
- Published
- Jun 2, 2026
- Updated
- Jun 2, 2026
- Reserved
- Jan 5, 2026
Description
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
Scores
| CVSS | Score | Vector | Source |
|---|---|---|---|
| 4.0 | 9.2 Critical | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N | VulnCheck |
| 3.1 | 9.8 Critical | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H | VulnCheck |
Weaknesses
- CWE-306 Missing Authentication for Critical Function
Affected Products
| Product | Versions |
|---|---|
SentinelSpacelabs Healthcare |
|
References
- https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdfvendor-advisorypatch
- https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinelproduct
- https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remotingthird-party-advisory
Every record here comes from the CVE Program’s CVE List. © 1999–2026 The MITRE Corporation. CVE is a registered trademark of The MITRE Corporation. The records are reproduced under the CVE Program terms of use.